Cart is empty
G-Series: Glass Slide-Based Arrays
Semi-Quantitative, Sandwich-based, Glass Slide Antibody Arrays |
|
|
Antibody arrays allow researchers to conduct rapid, accurate expression profiling of hundreds of cytokines, chemokines, growth factors, proteases, soluble receptors and other proteins from any biological fluid. RayBio® Glass Slide-based Antibody Arrays (G-Series) were introduced as a “second generation” array platform after the chemiluminescent membrane arrays (C-Series). The G-Series uses a 75mm x 25mm glass chip platform which delivers easier handling along with the use of much lower sample volumes. The glass slides also utilize a fluorescent signal readout, allowing a wider dynamic detection range than can be achieved by chemiluminescence. |
|
|
G-Series Antibody Arrays are particularly well-suited for projects with very limited sample volumes (as little as 10 μl per array) and large numbers of samples. Dozens of arrays can be processed per day by a single technician. Furthermore, the slide chamber assembly is compatible with many automated array processing systems, allowing high-throughput studies to be performed with very little hands-on time. For a comparison & overview of all our antibody array systems, click here. |
|
|
Features |
Research Applications |
|
|
|
|
|
How It Works
|
|
Similar to the C-Series (membrane-based) arrays, G-Series Antibody Arrays utilize the sandwich-ELISA design principle. In this assay, capture antibodies are printed in multiple identical arrays on a standard-sized histology slide. After a blocking step, samples are incubated with the arrays. Nonspecific proteins are then washed off, and the arrays are incubated with a cocktail of biotinylated detection antibodies, followed by a streptavidin-conjugated fluor. Signals are visualized using a fluorescence laser scanner. |
|
|
For a list of laser scanner specifications, click here. If you do not have access to a scanner, RayBiotech provides slide scanning and data analysis services |
||
Contents of kit
- RayBio® Cytokine Antibody Array glass slide (4 or 8 arrays per slide)
- Biotinylated Detection Antibodies
- Streptavidin-conjugated HiLytePlus™ 555 Fluor
- Blocking Buffer
- 20X Wash Buffer I
- 20X Wash Buffer II
- 2X Cell Lysis Buffer
- G-Series Antibody Array accessories*
- Manual
*Accessories include: 16-well incubation chamber with gasket, protective cover, snap-on sides, adhesive film